NEW! Shallow QC Sequencing for Fiber-seq
- 7 Articles
Yes! Before investing in deep sequencing, you can confirm that your Fiber-seq labeling worked as expected using Plasmidsaurus’s Fiber-seq for 6mA , developed in partnership with EpiCypher. First, assess the quality of your sample gDNA as...
The QC data is analyzed using a shorter version of the standard Fiber-seq data analysis workflow . Some familiarity with command-line tools is required. However, because the dataset is small (~100 Mb) the analysis can typically be run on a persona...
At this time, EpiCypher has validated the Fiber-seq QC for 6mA service for human and mouse samples only. If your organism’s genome is smaller than 3.1Gb, the 100Mb of data received should be sufficient to assess Fiber-seq success. If your organism...
Not necessarily. It is common to see differences between percent 6mA reported (and other QC metrics) from QC sequencing and full sequencing. Reasons include: The QC dataset is small, around 100 Mb, while full datasets are much larger. Diffe...
Plasmidsaurus needs at least 300ng of gDNA (15µL at ≥20ng/µL) per sample to perform this service.
EpiCypher can provide full Fiber-seq sequencing using our in-house PacBio Revio. This custom service can include preliminary data analysis and bioinformatics consultation to on-board Fiber-Seq data analysis capabilities in your lab. Reach out to ou...
Y es. EpiCypher routinely uses Plasmidsaurus Fiber-seq QC sequencing service even when the final sequencing will be performed on a PacBio Revio. You may observe small differences in reported percent 6mA between platforms, but this is expected due to...