Troubleshooting guidelines for CUT&RUN

Below we outline the main areas of concern and general considerations for troubleshooting CUTANA CUT&RUN assays. For ALL troubleshooting, we recommend reviewing the basic quality control metrics in Figure 1.

Troubleshooting low CUT&RUN DNA yields

If your experiment failed or if you have very low yields (< 5 ng), see this article and consider the following questions:

  • What is your cell/sample type? See Sample Prep section for specific recommendations.

  • Are Digitonin permeabilization conditions optimized for your cell type?

  • Have you confirmed sample prep and ConA bead binding?

  • Are you using the recommended 500,000 cells per reaction? Success from low cell numbers depends on antibody quality and target abundance. Review this article for guidance.

  • Have you included reactions with control antibodies and the K-MetStat Panel? These controls are crucial for troubleshooting CUT&RUN, as outlined here.

  • Confirm proper ConA bead storage conditions and quality. Are ConA beads brown and stored at 4˚C? ConA beads should NEVER be frozen.

  • Have reactions been mixed properly using a nutator? Have ConA beads become clumpy or dried out during the protocol?

Figure 1. Overview of CUTANA CUT&RUN quality control metrics.


Troubleshooting sequencing library yields and fragment distribution

Concern

Potential causes & troubleshooting approach

Low library yields, no enrichment in Bioanalyzer or TapeStation results

⚠ Low CUT&RUN yields, low inputs for library prep

⚠ Library prep technique

Adapter dimers

⚠ Self-ligation of sequencing adapters, preferentially amplified due to their small size


Troubleshooting CUT&RUN sequencing and results

Concern

Cause & troubleshooting approaches

Sequencing a low-concentration DNA library

⚠ If it is not possible to repeat library prep:

  • Use a Speedvac to increase the library concentration

  • Add as much of the library as possible to the sequencing pool

  • Deeper sequencing is recommended to capture read diversity

Background in open chromatin

⚠ Indicates over-digestion by MNase

  • Repeat assay with fresh buffers

  • Make sure MNase digestion is performed on ice

Experimental target shows high background and/or is indistinguishable from IgG negative control

⚠ Over-digestion by MNase, DNA damage, antibody failure. Use 500,000 cells per reaction; include reactions with control antibodies and the K-MetStat Panel. General tips:

  • Process cells quickly and resuspend in cold Antibody Buffer

  • Test multiple antibodies to experimental target

  • Ensure MNase digestion is incubated on ice for 2 hours

  • Keep adapter ligation reagents on ice during ligation setup