--- title: "What do adapter dimers look like in CUT&RUN and how do I avoid them?" slug: "what-do-adapter-dimers-look-like-in-cut-and-run" tags: ["cut and run"] updated: 2025-07-15T14:43:51Z published: 2025-07-15T14:43:51Z canonical: "support.epicypher.com/what-do-adapter-dimers-look-like-in-cut-and-run" --- > ## Documentation Index > Fetch the complete documentation index at: https://support.epicypher.com/llms.txt > Use this file to discover all available pages before exploring further. # What do adapter dimers look like in CUT&RUN and how do I avoid them? Adapter dimers are generated by self-ligation of sequencing adapters that are preferentially amplified due to their small size. Adapter dimer contamination appears as a peak at ~150 bp (below), and is caused by low input DNA, inefficient adapter ligation, and/or using excess beads during library purification. Adapter dimers should represent no more than 5% of the total 200-700 bp fragment yield as determined on a Bioanalyzer/Tapestation trace (i.e. % Integrated in trace is less than 5%). To minimize adapter dimers, keep adapter ligation reagents on ice during ligation setup. If adapter dimers are >5% of your library, they should be removed using the bead:DNA ratios outlined in the Application Notes for our CUTANAâ„¢ DNA Purification Beads ([EpiCypher 21-1407](http://www.epicypher.com/product/cutana-dna-purification-beads/)) ![adapter dimers@2x](https://cdn.document360.io/1fc69f91-cf1e-45ab-bc51-4063eb80e37c/Images/Documentation/adapter%20dimers@2x.png) **Figure 1.** Example TapeStation trace from CUT&RUN H3K27me3 library containing an adapter dimer peak (~150 bp, red arrow) and expected library peak (~300 bp, blue arrow). Red lines denote the 200-700 bp range, used to determine library concentration.